%0 Journal Article %T 动态稀释校准仪对VOCs自动监测系统的影响 %T Influence of Dynamic Dilution Calibrator on VOCs Automatic Monitoring System %A 田新愿,张健,饶正勇,秦兴章,晁良俊 %A TIAN,Xin yuan %A ZHANG,Jian %A RAO,Zheng yong %A QIN,Xing zhang %A CHAO,Liang jun %J 环境监控与预警 %@ 1674-6732 %V 11 %N 4 %D 2019 %P 28-31 %K 动态稀释校准仪;稀释比;稀释流量;自动监测系统;挥发性有机物 %K Dynamic dilution calibrator; Dilution ratio; Dilution flow; Automatic monitoring system;VOCs %X 基于环境空气VOCs自动监测系统中动态稀释校准仪在相同稀释比和不同稀释流量条件下,对仪器测试结果误差的影响进行研究,指出了当前环境空气VOCs自动监测普遍使用的校准仪的不足之处。选取臭氧前体物(PAMS) 57种有机化合物,以不同稀释总流量,分别测试0.5和1.0 nmol/mol的平均检出限以及平均回测偏差,结果显示,在输出流量>1 000 mL/min时,平均检出限分别在0.02~0.17和0.01~0.10 nmol/mol;平均回测偏差在-12%~26%和-28%~10%。当总输出流量>1 000 mL/min时,多点曲线各浓度点重复性<5%。 %X Based on the study of the influence of the dynamic dilution calibrator in the ambient atmospheric VOCs automatic monitoring system on the error of the instrument test results under the same dilution ratio and different dilution flow conditions, the shortcomings of the current calibrator commonly used in environmental VOCs automatic monitoring are pointed out. 57 kinds of organic compounds of PAMS were selected, the average detection limits and the average back-test deviation of 0.5 nmol/mol and 1.0 nmol/mol were tested at different dilution total flow rates. The results showed that the average detection limits were 0.02~0.17 nmol/mol and 0.01~0.10 nmol/mol; the average back-test deviation was -12%~26% and -28%~10%, when the total flow rate was ≥1000 mL/min. Meanwhile, when the total flow rate is above 1000 mL/min, the repeatability of each concentration point of the multi-point curve is within 5%. %R %U http://www.hjjkyyj.com/hjjkyyj/ch/reader/view_abstract.aspx %1 JIS Version 3.0.0