空气自动监测系统运行管理故障处理
引用本文:王,卫.空气自动监测系统运行管理故障处理[J].环境监控与预警,2011,3(1):23-24.
作者姓名:  
作者单位:宿迁市环境监测中心站,江苏,宿迁,223800
摘    要:结合宿迁市空气质量自动监测系统运行实践,解析API点式仪器SO2、NOX分析仪校标不稳、测值偏低,BAM-1020BETA射线颗粒物监测仪纸带易破,数据采集系统线路不畅、数据不完整等常见故障,提出相应的解决方法和预防对策.

关 键 词:环境空气  自动监测  运行  故障
收稿时间:2/9/2010 12:00:00 AM
修稿时间:2010/2/21 0:00:00

The Fault and Disposal of the Operation and Management of Air Automatic Monitoring
WANG Wei.The Fault and Disposal of the Operation and Management of Air Automatic Monitoring[J].Environmental Monitoring and Forewarning,2011,3(1):23-24.
Authors:WANG Wei
Affiliation:WANG Wei(Suqian Environmental Monitoring Central Station,Suqian,Jiangsu 223800,China)
Abstract:With the operation practices of the air quality automatic monitoring system in Suqian, the problems of inaccurate calibration of SO2,NOX analysis apparatus and underestimated observed value were analyzed. BAM-1020 BETA-ray monitor tape was easily broken. Poor line in data acquisition system, missing data and other common faults were also discussed. Finally, appropriate solutions were proposed to these problems.
Keywords:ambient air  automatic monitoring  operation  fault